People with affected hip replacements should expect to be
contacted by their hospital for a clinical assessment following a
review by the Medicines and Healthcare products Regulatory Agency
(MHRA) which found increased safety risks for patients with
specific Profemur cobalt chrome modular neck hip replacements.
These are:
-
All hip replacement constructs using Profemur cobalt chrome
(CoCr) modular neck components;
-
All Profemur Xm hip replacement stems (cobalt chrome) used
with Profemur titanium or cobalt chrome modular neck
components.
While most patients implanted with these hip replacements
experience good outcomes and are unlikely to develop serious
problems, a small number may be affected by soft tissue reactions
or other metal-wear effects, which could lead to the need for
revision surgery.
UK trusts and hospitals are being advised to identify and contact
patients implanted with affected devices to offer them a
follow-up assessment as soon as practically possible, based on
their individual risk level. Patients do not need to contact
their surgeon or hospital directly.
This advice does not apply to all Profemur branded products, as
the brand covers a wide range of products.
The affected devices were used in the UK from 2009 until January
2025 and the available data suggests fewer than 2,000 patients
currently have these particular hip replacements implanted. These
devices are no longer available in the UK.
For context, more than 100,000 hip replacements of all types are
performed in the UK each year.
Dr Alison Cave, MHRA Chief Safety Officer,
said:
“Our review identified higher than expected risks associated with
Profemur cobalt chrome modular neck hip replacements.
“We have developed specific follow-up advice to help ensure
patients with these devices receive appropriate care. If you have
an affected hip replacement device implanted, you should expect
to be contacted by your surgeon or implanting hospital. You don't
need to contact them directly.
“Anyone experiencing new or unexpected symptoms related to any
implant should speak to their surgeon or the hospital where their
surgery was performed. Please report it to the MHRA via the
Yellow Card scheme or the Incident Reporting & Investigation
Centre if you are in Scotland.”
For more information, visit: Profemur Cobalt Chrome
Modular Neck Hip Replacements (DSI/2025/005).
Notes to editors
-
The MHRA review found that the affected devices carry a
higher risk of wear, corrosion, and fracture – impacting
around 6 in every 1,000 patients, compared to fewer than 1 in
10,000 with alternative hip replacement designs.
-
This advice does not apply to all Profemur branded products,
as the brand covers a wide range of products. Only the
product codes included in the DSI are
affected.